S = Sign. A minus sign (
−
) indicates a negative number. A blank indicates a positive
number.
•
X.XXX = The 4-digit mantissa.
•
E = The exponent indicator.
•
YY = The 2-digit exponent.
Procedure
1.
Switch from decimal notation to exponential notation.
a. Activate
Select
as required until the rightmost digit is flashing.
b. Activate
Scroll
until
E
is displayed.
c. Activate
Select
.
Tip
If you have modified the value in decimal notation without saving the changes to transmitter
memory, the changes will be lost when you switch to exponential notation. Save the decimal
value before switching to exponential notation.
2.
Enter the exponent.
The first character may be a minus sign or any digit between 0 and 3. The second
character may be any digit between 0 and 9.
a. Activate
Select
to move the cursor to the rightmost character on the display.
b. Activate
Scroll
until the desired character is displayed.
c. Activate
Select
to move the cursor one position to the left.
d. Activate
Scroll
until the desired character is displayed.
3.
Enter the mantissa.
The mantissa must be a 4-digit value with a precision of 3 (that is, all values between
0.000 and 9.999).
a. Activate
Select
to move the cursor to the rightmost digit in the mantissa.
b. Activate
Scroll
until the desired character is displayed.
c. Activate
Select
to move the cursor one digit to the left.
d. Activate
Scroll
until the desired character is displayed.
e. Activate
Select
to move the cursor one digit to the left.
f. Activate
Scroll
until the desired character is displayed.
g. Activate
Select
to move the cursor one digit to the left.
h. Activate
Scroll
until the desired character is displayed.
4.
Enter the sign.
a. Activate
Select
to move the cursor one digit to the left.
b. Activate
Scroll
until the desired character is displayed.
For positive numbers, select a blank space.
Using the transmitter display
Configuration and Use Manual
175
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...