10.19
Check
mA Output Fault Action
mA Output Fault Action
controls the behavior of the mA output if the transmitter encounters
an internal fault condition. If the mA output is reporting a constant value below 4 mA or
above 20 mA, the transmitter may be in a fault condition.
1.
Check the status alerts for active fault conditions.
2.
If there are active fault conditions, the transmitter is performing correctly. If you
want to change its behavior, consider the following options:
•
Change the setting of
mA Output Fault Action
.
•
For the relevant status alerts, change the setting of
Alert Severity
to
Ignore
.
Restriction
For some status alerts,
Alert Severity
is not configurable.
3.
If there are no active fault conditions, continue troubleshooting.
Related information
10.20
Check for radio frequency interference (RFI)
The meter's TPS output or discrete output can be affected by radio frequency interference
(RFI). Possible sources of RFI include a source of radio emissions, or a large transformer,
pump, or motor that can generate a strong electromagnetic field. Several methods to
reduce RFI are available. Use one or more of the following suggestions, as appropriate to
your installation.
Procedure
•
Use shielded cable between the output and the receiving device.
-
Terminate the shielding at the receiving device. If this is impossible, terminate
the shielding at the cable gland or conduit fitting.
-
Do not terminate the shielding inside the wiring compartment.
-
360-degree termination of shielding is unnecessary.
•
Eliminate the RFI source.
•
Move the meter.
10.21
Check
Flow Direction
If
Flow Direction
is set inappropriately for your process, the meter may report velocity data
that is not appropriate for your requirements.
Troubleshooting
Configuration and Use Manual
161
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...