Method
Description
Setup
Polling
The meter polls an external de-
vice for temperature data. This
data will be available in addi-
tion to the internal tempera-
ture data.
a. Choose
Configure
>
Manual Setup
>
Measurements
>
External Inputs
>
Temperature
.
b. Set
External Temperature
to
Enable
.
c. Choose
Configure
>
Manual Setup
>
Inputs/Outputs
>
External Device
Polling
.
d. Choose an unused polling slot.
e. Set
Poll Control
to
Poll as Primary
or
Poll as Secondary
.
Option
Description
Poll as Primary
No other HART masters will be on the
network. The Field Communicator is not
a HART master.
Poll as Secondary
Other HART masters will be on the net-
work. The Field Communicator is not a
HART master.
f. Set
External Device Tag
to the HART tag of the external tem-
perature device.
g. Set
Polled Variable
to
Temperature
.
Digital communica-
tions
A host writes temperature data
to the meter at appropriate in-
tervals. This data will be availa-
ble in addition to the internal
temperature data.
a. Choose
Configure
>
Manual Setup
>
Measurements
>
External Inputs
>
Temperature
.
b. Set
External Temperature
to
Enable
.
c. Perform the necessary host programming and communica-
tions setup to write temperature data to the transmitter at
appropriate intervals.
2.
Choose the method to be used to supply pressure data, and perform the required
setup.
Configure process measurement
Configuration and Use Manual
55
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...