9
Measurement support
Topics covered in this chapter:
•
Perform the Known Density Verification procedure
•
Adjust density measurement with
•
Perform density offset calibration
•
Adjust temperature measurement with
•
Adjust concentration measurement with
•
Adjust concentration measurement with
•
Set up user‐defined calculations
9.1
Perform the Known Density Verification
procedure
The Known Density Verification procedure is used to verify that the meter's current
operation matches the factory baseline. If the meter passes the test, the meter does not
have any physical problems such as denting, bending, twisting, erosion, or corrosion.
•
Perform the Known Density Verification procedure using the display
(Section 9.1.1)
•
Perform the Known Density Verification procedure using ProLink III
(Section 9.1.2)
•
Perform the Known Density Verification procedure using the Field Communicator
(Section 9.1.3)
9.1.1
Perform the Known Density Verification procedure using
the display
The Known Density Verification procedure is used to verify that the meter's current
operation matches the factory baseline. If the meter passes the test, the meter does not
have any physical problems such as denting, bending, twisting, erosion, or corrosion.
Prerequisites
Power down the meter, remove the meter from the process, and place it in a protected
stable environment.
Ensure that the sensor tubes are clean and dry.
Minimize variation in ambient temperature.
Eliminate or minimize vibration.
Power up the meter.
Measurement support
Configuration and Use Manual
119
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...