Note
Temperature offset is applied only to temperature data from the on-board temperature sensors
(RTD). External temperature data is not adjusted.
Prerequisites
You will need an external temperature measurement method that is highly accurate.
Ensure that your process is stable during the sampling procedure. Minimize variation in
density, temperature, flow rate, and fluid composition. Minimize aeration.
Procedure
1.
Take a temperature reading from the meter.
2.
Immediately after the previous step, take a sample from a location as close to the
meter as possible.
3.
Using the external measurement method, measure the temperature of the sample.
4.
Subtract the measured temperature value from the externally measured value.
5.
Set
Temperature Offset
to the result.
•
Using the display: Not available
•
Using ProLink III:
Device Tools
>
Configuration
>
Process Measurement
>
Line Temperature
>
Temperature Offset
•
Using the Field Communicator: Not available
The default value is 0. The range is unlimited.
9.5
Adjust concentration measurement with
Trim
Offset
Trim Offset
adjusts the meter's concentration measurement to match a reference value.
Tip
You can adjust concentration measurement by applying the trim offset only, or by applying both the
trim offset and the trim slope. For most applications, the trim offset is sufficient.
Prerequisites
You must be able to take a sample of your process fluid and obtain a laboratory
concentration value at line density and line temperature.
Procedure
1.
Take a concentration reading from the meter, and record line density and line
temperature.
2.
Take a sample of the process fluid and obtain a laboratory value for concentration at
line density and line temperature, in the units used by the meter.
Measurement support
128
Micro Motion
®
Compact Density Meters (CDM)
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...