Assembly and preparation
94
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Locking/unlocking the M540
You can lock the M540 in the M500 to prevent any-
one from undocking it.
To lock the M540 into place
1
Push the locking tab (D) of the M500 towards
the front. This prevents you from undocking the
M540. If you push the locking tab back, you can
undock the M540 again.
2
Insert the 2 mm hex wrench tool (A) into the
middle hole (B) on the locking tab and turn it
clockwise to the locked position .The
locking
tab is fixed and you cannot remove the M540
unless you unlock it again using the hex wrench
tool.
A
Hex wrench tool (2 mm)
B
Center hole on locking tab for locking/unlocking
the M540
C
Release buttons for undocking the M540
D
Locking tab
To unlock the M540
1
Insert the 2 mm hex wrench tool (A) into the
middle hole (B) on the locking tab and turn it
counterclockwise to the unlocked position
.
2
Push the locking tab (D) back to unlock the re-
lease buttons (C) on the M500 to undock the
M540.
307
Top view of the M500
A
C
C
B
D