External Device – Neuromuscular transmission (NMT)
392
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
NMT parameter functions and settings
Selection
Available settings
Description
Display
temperature
– On
–
Off
(Default)
Controls the temperature
display in the NMT parameter
field.
Stim current
0 to 60 mA (no default)
Displays the stimulation current
of the NMT device.
Stim current
mode
Auto
The NMT device establishes a
supramaximal current during
the first measurement and uses
it for subsequent
measurements.
Manual
The stimulation current was
selected manually.
Sensitivity (1-512)
Between 1 and 512 (no default)
Displays the sensitivity of the
NMT device.
Pulse width
200 or 300 µs
Determines the width of the
pulse.
Prints the current NMT settings and up to 500 of the most recent measurements.
No alarm
signaling for this
device.
Informational message that no optical or acoustic alarm signals are available on
the Cockpit.