Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
465
System configuration
Alarm volume off
– Yes
–
No
(default)
–
Yes
– the acoustic alarm signal for any
active alarm is deactivated, the
Audio
alarms off
message and the
symbol
appear in the header bar
.
–
No
– the acoustic alarm signals for any
active alarm are not affected when you
press the
Code
button.
All alarms off
– On
–
Off
(default)
–
On
–
the following happens when you select
the
Code
button.
– All audible and optical alarm signals are
deactivated at the Cockpit and at the
ICS.
– The message area flashes red with the
message
All alarms off
. The
symbol appears in the header bar.
– The
All alarms paused
setting is set to
No timeout
– if the selection
All alarms off reminder
is set to
On
, an alarm tone sounds every
30 seconds to indicate that the alarms
were disabled.
Selecting the
Code
button again causes
any existing alarm conditions to be
annunciated immediately.
–
Off
– any existing alarm is still annunciated
optically and acoustically when you select
the
Code
button.
Selection
Available settings
Description