Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
375
Scio Monitoring
Manual agent identification
Manual agent identification is available only for gas
analyzers without automatic agent identification:
Scio Four and Scio Four Oxi.
To configure manual agent identification, refer to
"Agent parameter setup functions" on page 365.
Automatic agent identification
Automatic agent identification setup is available
only for the following gas analyzers: Scio Four Plus
and Scio Four Oxi plus.
These gas analyzers automatically identify up to
two anesthetic agents, even in mixtures.
If the gas analyzer has not yet identified or cannot
identify an agent, or has detected a mixture of three
or more anesthetic agents (for example, due to too
low agent concentrations, a leaking vaporizer, or
traces of disinfectants), the agent parameter field is
blank and the Agent label displays
Agent?
Mixed Agent
When the gas analyzer detects a mixture of two
anesthetic agents, the displayed real-time
waveform is the primary agent concentration. The
color of the waveform and the first agent label in the
Agent
parameter field represents the agent with
the highest expiratory xMAC value (primary agent).
The following label indicates the SEV agent with
the highest expiratory xMAC value:
This label switches to that of the second
administered agent when its expiratory xMAC value
exceeds that of the first agent:
WARNING
Risk due to inaccurate gas measurement val-
ues
Use care when selecting the agent manually.
Measurements are inaccurate if the wrong
agent is selected.
WARNING
Risk due to inaccurate gas measurement val-
ues
Measurements using a gas analyzer without
automatic agent recognition are inaccurate if
anesthetic gases are mixed.
NOTE
The measurement of mixed agent is available
only for gas analyzers with automatic agent rec-
ognition: Scio Four plus and Scio Four Oxi plus.
Sev
Des
Sev
Des