Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
441
External devices – Evita 2D, Evita 4, Evita XL (Medibus)
CO
2
parameter setup functions
See page 439 for information on how to access this
page.
Selection
Available settings
Description
CO
2
Scale
– 0 to 40 (default), 0 to 60, 0 to 80,
0 to 100 mmHg
– 0.0 to 5.0 (default), 0.0 to 8.0, 0.0 to 12.0,
0.0 to 16.0 kPa
– 0 to 5 (default), 0 to 8, 0 to 12, 0 to 16%
Determines the scale of the dis-
played CO
2
waveform.
Atm. pressure
570 to 800 mmHg
Determines the ambient pres-
sure setting.
Color
Red, green, blue, yellow (default), light blue,
purple, orange, white.
Determines the color of the CO
2
parameter field and waveform.