Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
537
Troubleshooting
Temperature
Priority
Message
Parameter field
Problem
Solution
!!
T1a
> (alarm limit)
T1b
> (alarm limit)
Ta
> (alarm limit)
Tb
> (alarm limit)
T
> (alarm limit)
∆
T
> (alarm limit)
∆
T1
> (alarm limit)
T1a
< (alarm limit)
T1b
< (alarm limit)
Ta
< (alarm limit)
Tb
< (alarm limit))
T
< (alarm limit)
∆
T
< (alarm limit)
∆
T1
< (alarm limit)
Parameter value
The parameter
value is
above/below the
upper/lower
alarm limits.
– Check the
patient and
treat if
necessary.
– Change the
alarm limits.
!!
%0 out of range
high
1)
The parameter value
is replaced by +++
The parameter
value is
above/below the
measurement
range of the
monitor.
– Check the
patient and
treat if
necessary.
– Check the
equipment and
replace, if
necessary.
!!
%0 out of range low
1)
The parameter value
is replaced
by - - -
!
Cannot derive %0
2)
The parameter value
is replaced by ***
The cable is
either faulty or
unplugged.
– Check the
equipment and
replace it if
necessary.
– Connect the
second
temperature
sensor.
!
%0 H/W failure
1), 2)
The parameter value
is replaced by ***
The hardware
reference values
do not meet the
specified
tolerance.
Contact
DrägerService
®
.
!
%0 Unplugged
1), 2)
The parameter value
is replaced by ***
The temperature
sensor is
unplugged.
Reapply the
temperature
sensor.