![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 253](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773253.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
253
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
Standard parameter set
The Infinity MCable – Masimo SET and the Masimo
rainbow SET MCable always support the following
parameters:
– Functional oxygen saturation (SpO2). The unit
of measurement is%.
– Pulse rate (PLS). The unit of measurement is
beats/min.
– Perfusion index (PI) which indicates the arterial
pulse signal strength. The measurement is 0-1.
Expanded parameter set
In addition to the above standard parameters, the
Masimo rainbow SET MCable provides the
following additional optional parameters:
– Total hemoglobin (SpHb) measures the total
hemoglobin levels in arterial or venous blood.
The unit of measurement is selectable (see
page 472).
– Total oxygen content (SpOC) measures the
total blood oxygen content; this value is
calculated from the SpHb and the SpO2 values.
The unit of measurement is mL/dL.
– Pleth variability index (PVI) measures
peripheral perfusion changes secondary to
respiration or the PI amplitude over a
respiration. PVI may be closely related to
intrathoracic pressure changes, circulating
blood volume and vascular tone. The unit of
measurement is%.
– Carboxyhemoglobin saturation (SpCO)
measures the amount of carbon monoxide that
is bound to hemoglobin. The unit of
measurement is%.
– Methemoglobin saturation (SpMet) measures
the methemoglobin concentration in arterial
blood. The unit of measurement is%.
Various sensors are available for the Masimo
rainbow SET MCable. The availability of the
parameters depends on the selected sensor type.
Each sensor provides certain parameters which
must also be activated on the Masimo rainbow SET
MCable.
– CO SpO2 sensor; this type of sensor provides
the following parameters: SpO2, PLS, PI,
SpCO, SpMet, PVI.
– M-LNCS sensor; this type of sensor provides
the following parameters: SpO2, PLS, PI.
– Hb sensor; this type of sensor provides the
following parameters: SpO2, PLS, PI, SpHb,
SpOC, SpMet, PVI.
The following figure shows the multi-color band
which appears on the side of the Masimo rainbow
SET MCable (see page 256 for more information).
If you connect a sensor but the parameter is not
activated on the MCables, the parameter label
appears in the parameter field without a value.
NOTE
A color band on the Masimo rainbow SET MCable
indicates which parameters are activated on the
MCable. If an MCable does not have a label, the
supported parameters are by default SpO2, PLS,
and PI.
478