System configuration
470
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Reports setup –
Reports setup
page
The following table lists the available settings of the
Reports setup
page. To access this page, see
page 467.
Biomed setup
This section describes several pages accessible
only to authorized personnel. All Biomed pages are
password protected.
To access the biomed pages
1
Select
System setup...
on the main menu bar.
2
Select the
Biomed
tab.
3
Enter the password and select
Enter
.
4
Select one of the following tabs:
– Country
– Units of measure
– Patient monitor
–
Name service
– Network setup
(select either the
Infinity
or
Hospital
tab)
– Printer setup
– Recorder setup
– Service
– IT setup
(select the desired tab such as
Web browser
,
Innovian
,
and so on).
NOTE
Reports printed on a laser printer use the settings
in the
Reports setup
page not the settings de-
fined in the trend setup pages.
Selection
Available settings
Description
Waveform delay
[s]
6, 10 (default), 15 s
Determines the amount of delay (pre-event)
data included in a timed strip report. Delay data
refers to data that originated
before
the report
was initiated. A marker on the report marks
where the delay data ends and the real-time
data starts.
Waveform
duration [s]
10, 20 s (default)
Determines the length of a strip report.
Trend duration
[hr]
1, 2, 4, 8, 12, 24 (default), 48,
72, 96 hr
Determines the graphical trend interval on the
graphical trend report.
Table interval
[min]
1, 5, 10, 15 (default), 30,
60 min
Determines the tabular trend interval on the
tabular trend report.
WARNING
Do not service the Cockpit while monitoring a
patient.