Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
353
Microstream CO
2
monitoring
Using the CO
2
dialog box
Setup for CO
2
parameters takes place in the
CO
2
dialog. This dialog contains the following tabs:
– Mainstream
sets associated CO2 parameters
– Microstream
sets associated CO
2
parameters
–
Calibration check
performs a
Microstream MCable calibration. For
information about the Calibration Check tab,
see "Calibration Check" on page 355.
If the Cockpit is not set up for CO
2
monitoring, then
the
Mainstream
tab also displays in the
CO
2
dialog.
To access the Microstream settings
If the
CO
2
parameter displays on the Cockpit:
1
Touch the
CO
2
parameter field.
2
Touch the
Microstream
tab.
Or, if the
CO
2
parameter field is not displayed:
Select
Sensor parameters...
from the main
menu bar >
Microstream
tab to access the
Microstream
page. If you cannot see the tab,
select the following two symbols located in the
upper right corner of the dialog:
symbol and
the display filter
button.
CO
2
limits
Setup functions for CO
2
parameters take place in
the
Microstream
tab within the
CO
2
dialog.
The
Microstream
tab displays fields on the upper
left side of the window for
etCO2
,
inCO2,
and
RRc
settings.
etCO2
and
RRc
allow adjustment of the upper and
lower alarm limits, use of the
Archive
feature, and
use of
Auto set.
For inspiratory CO
2
, only the
upper limit
and
Archive
can be used.
NOTE
CO
2
limits settings apply to both Mainstream and
Microstream.