Operating concept
50
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
3
Select the
Start/Standby...
button on the main
menu bar.
4
Select the
Transfer
tab (if not already
selected). The
Transfer
page (see page 54)
lists all of the devices in the currently selected
care area who are in standby mode.
5
Use the care area selection arrow (B) on the
Transfer
page to select the care area from the
list in which the source device is located.
6
Select the source device in the
Device name
column (D).
7
Select the
Start transfer
button (E).
8
Press the rotary knob. A
Confirm transfer
dialog appears with the following message:
The
network transfer will delete existing patient
data on this device.
9
Select one of the following buttons:
–
Cancel –
to prevent the data transfer and
return to the
Transfer
page.
–
Transfer –
to discharge the patient from the
Cockpit and initiate the data transfer. During
the transfer, the Cockpit and M540 display
the message
Transferring data...
A successful transfer results in the following:
– The Cockpit returns to the main screen.
– The message
Transfer Complete
appears in
the Cockpit header.
A failed transfer results in the following:
– The Cockpit returns to the
Transfer
page.
– The message
Transfer of data incomplete
displays in the network message header of the
Cockpit.
– The M540 returns to the last monitoring screen
before the transfer started.
CAUTION
To prevent mixing the data of two patients, first
discharge the patient admitted at the destination
Cockpit. If the patient is not discharged, the new
data is appended to the existing data stored at the
destination Cockpit.
CAUTION
Do not touch the screen or undock the M540
during the transfer, or a network failure will occur,
and the transfer will not be completed.