Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
215
ECG, arrhythmia, and ST segment
Cable type
(TruST is only
available with a 6-
wire lead set)
–
Auto detect
(default)
– 3-, 5-, 6-electrodes, and 12 leads
When using the ECG extension cable, the
system always assumes the cable is a 6-wire
lead set.
When set to
Auto detect
, this
feature detects the number of
connected lead wires
automatically. If auto detect
mode does not detect the
connected lead set, it allows
you to select the cable type
manually. “12” denotes a
combination of a 6-wire lead
set and 4- wire lead set for 12-
lead monitoring.
ARR Processing
– ECG1
–
ECG1 & 2
(default)
The
ECG1 & 2
selection is not available if the
neonatal patient category is selected.
ECG1
setting – arrhythmia
processing occurs only on the
lead displayed in waveform
channel 1.
ECG1 & 2
setting –
arrhythmia processing occurs
on the leads displayed in the
waveform channels 1 and 2.
Size all ECG
[mV/cm]
0.25, 0.5, 1 (default), 2, 4, 8 mV/cm
Sets the amplitude of all
displayed ECG leads.
QRS threshold
–
Normal
(default)
– Low
WARNING
Risk of inaccurate HR value
If the QRS setting is set to Low in the pres-
ence of HR artifact, the associated HR value
may be inaccurate.
To avoid an inaccurate HR value, it is recom-
mended to set the QRS threshold setting to
Normal.
This function is only available
for adult and pediatric
patients.
Normal
– detects QRS
complexes of normal
amplitude.
Low
– detects QRS
complexes of low amplitude.
Resp. monitoring
–
On
(default for neonate)
–
Off
(default for adult/pediatric)
Activates/deactivates
respiration monitoring.
Show all
page
This page displays all available leads (up to 12).
Selection
Available settings
Description