Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
155
Alarms
IBP M pediatric
Increment:
1 mmHg or
0.1 kPa
On
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
– 80 mmHg
(10.7 kPa)
for GP1 M to
GP4 M,
ART M, LV M
– 20 mmHg
(2.7 kPa) for
LA, ICP, CVP
– 17 mmHg
(2.3 kPa) for
PA M, RV M
– 12 mmHg
(1.6 kPa) for
RA
– 50 mmHg
(6.7 kPa)
for GP1 M
to
GP4 M,
ART M
– 40 mmHg
(5.3 kPa)
for LV M
– 7 mmHg
(0.9 kPa)
for PA M,
RV M
– 2 mmHg
(0.3 kPa)
for RA, ICP
– 0 mmHg
(0.0 kPa)
for LA,
CVP
Off
IBP M neonate
Increment:
1 mmHg
or 0.1 kPa
On
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
– 85 mmHg
(11.3 kPa) for
GP1 M to
GP4 M, ART
M,
– 80 mmHg
(10.7 kPa)
for
LV M
– 20 mmHg
(2.7 kPa) for
LA, ICP, CVP
– 17 mmHg
(2.3 kPa) for
PA M, RV M
– 12 mmHg
(1.6 kPa) for
RA
– 40 mmHg
(5.3 kPa)
for GP1 M
to
GP4 M,
ART M,
LV M
– 7 mmHg
(0.9 kPa)
for PA M,
RV M
– 2 mmHg
(0.3 kPa)
for RA, ICP
– 0 mmHg
(0.0 kPa)
for LA,
CVP
Off
Parameter Alarm
default
status
Alarm limit range
Upper limit
defaults
Lower limit
defaults
Archive
default
setting