Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
519
Troubleshooting
!!!
Ventricular
fibrillation
VF
The reported
arrhythmia was
detected
Check the patient and
treat if necessary.
Rest ECG messages
None
ECG Collecting
waveforms
Parameter value
Rest ECG was
initiated
Instruct the patient to
lie still.
None
ECG busy
Parameter value
The central station
is already
processing a
report.
Wait a few minutes
before requesting the
report again.
None
ECG cannot
connect
Parameter value
Connection to
central station is
not possible.
– Check that the
patient is
admitted at the
central station.
– Check that the
central station
has the Rest ECG
option activated.
None
ECG report
complete
Parameter value
The Rest ECG
report has been
printed
Informational
message – no action
required.
None
%0 comm failure
1)
Parameter value
The external
device is not
available.
Check the
configuration at the
central station.
None
Sending ECG data
Parameter value
Informational
message.
Informational
message – no action
required.
1)
%0 is a placeholder for the parameter label HR or ECG.
Priority
Message
Parameter field
Problem
Solution
NOTE
RRi
and 12-lead ECG monitoring are unavailable
when the M540 is in OR mode and the ECG filter
is set to
Monitor
.