![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 384](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773384.webp)
External Device – Bispectral index (BIS)
384
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Accessing the BIS settings
Select the BIS parameter field to select the
BIS
page directly.
or
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the
BIS
tab to access the
BIS
page.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog:
symbol and the display filter
button.
3
Select the
Settings
and
Show all
tab.
BIS parameter setup functions
All BIS setup functions take place in the
BIS
page.
Selection
Available settings
Description
Settings
page
Scale[uV]
5, 10, 25, 50, 100 (default), 250 µV
Selects the scale of the EEG T
waveform.
Smoothing rate
[s]
Informational data only
Not applicable
BIS secondary
parameter
BSR, BCT (default), SEF, PWR
Selects a secondary parameter
and assigns it to the BIS
parameter field (see page 381).
No alarm
signaling for this
device.
Informational message that no optical or
acoustic alarm signals are available on the
the Cockpit
Not applicable
BIS
Show all
page
This page displays all supported BIS parameters, settings, labels and the units of measure where
appropriate.