Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
547
Troubleshooting
Microstream CO
2
Calibration and maintenance
Priority
Message
Cause
Remedy
None
CO
2
calibration check
failed
The Microstream MCable
calibration procedure failed.
Contact DrägerService
®
.
None
CO
2
calibration check
successful
The Microstream MCable
calibration procedure was
successful.
Informational message – no
action required.
None
CO
2
calibration check in
progress
The Microstream MCable
calibration procedure is in
progress.
Informational message – no
action required.
None
CO
2
calibration required
The Microstream MCable
calibration procedure is due.
Contact DrägerService
®
.
None
CO
2
MCable:
Maintenance is due
Maintenance for
Microstream MCable is due.
Contact DrägerService
®
.
None
%0 zeroing failed
Resetting the Microstream
MCable to zero has failed
the standard three attempts.
Contact DrägerService
®
.
None
%0 zeroing in progress
The Microstream MCable is
being reset to zero.
Informational message – no
action required.
None
CO
2
zeroing succesful
The Microstream MCable is
successfully reset to zero.
Informational message -- no
action required.