System configuration
466
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Alarm setup – configuring M540 settings
The
M540 settings
page configures certain M540
settings. These setting will automatically update the
M540
settings when the M540 is docked. To access
this page, see page 458.
Selection
Available settings
Description
Transport alarm
volume
– 50% (default) to 100%
(in increments of 10%)
Determines the speaker volume of the M540 at
the Cockpit while the M540 is undocked.
The Cockpit setting adjusts the
Transport
alarm volume
setting to match the M540 when
it is undocked.
The
Transport alarm volume
setting is tied to
the minimum alarm volume setting (see
page 461). If the minimum alarm volume is set
to a higher volume than the selected
Transport
alarm volume
setting, the
Transport alarm
volume
setting is adjusted to the higher setting.
If the minimum alarm setting is set to a lower
setting than the current
Transport alarm
volume
setting, the setting does not change.
Transport pulse
tone volume
– Off
(default)
– 5%
– 10% to 100%
(in increments of 10%)
Determines the pulse tone volume of the M540
while the device is on transport.
Keep device label
– Yes
(default)
– No
–
Yes
– The M540 retains the product label of
the Cockpit when it undocks.
–
No –
the M540 retains the product label
configured in the M540 wireless menu.