Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
587
Technical data
Infinity PS250 power supply
CAUTION
The medical device must only be used with
software that is tested and approved by Dräger.
Any modifications of the operating system
settings can impair operating safety.
Responsibility for any such modifications lies with
the operating organization.
Physical specifications
Dimensions (W x D x H)
27.76 x 11.68 x 34.59 cm (10.93 x 4.60 x 13.62 in)
Weight
5.5 kg (12 lbs)
Cooling
Convection
Connections
– AC power
– RS232
– Alarm output
– System cable
– Infinity network (Ethernet)
– Potential equalization connector
Environmental specifications
Humidity (non-condensing)
Operating: 20 to 95%
Storage/Transport: 20 to 95%
Temperature
Operating: 0 to 40°C (32 to 104°F)
Storage: –20 to +60°C (–4 to +140°F)
The PS250 has an operating temperature range of
0 to 40°C (32 to 104°F). When the battery is being charged
at ambient temperatures below 5°C (41°F), the yellow LED
on the PS250 may light up.
Ambient pressure
Operating: 485 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Protection against ingress of water
IPX1 according to IEC 60529 – protected against harmful
effects of dripping water when mounted vertically with the
connectors facing down.