Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Connecting the CO
2
sensor
Before connecting any CO
2
hardware, make sure
the airway adapter in use matches the airway
adapter setting of the Cockpit. For example, do not
use a disposable airway adapter if the Cockpit is
configured for a reusable airway adapter (and vice
versa). Not aligning the adapter with the
configuration setting at the Cockpit compromises
the displayed CO
2
value.
The CO
2
cable connects directly to the M540.
A
M540 CO
2
port
B
CO
2
sensor cable connector
C
Mainstream airway adapter
D
Y-piece
E
Airway adapter
F
Endotracheal tube adapter
To connect the hardware
The IACS is only compatible with the sensors
6871950 revision 5 or higher. Previous revisions
are not compatible.
1
Connect the CO
2
sensor cable connector (B) to
the yellow CO
2
port (A) on the M540.
2
Select a suitable Mainstream airway adapter
(C) whose windows are clean and dry (replace
the adapter if necessary).
3
Insert the airway adapter (E) between the
endotracheal tube adapter (F) and the ventilator
Y-piece (D).
4
Snap the Mainstream airway adapter (C) firmly
into the airway adapter and make sure that the
cable is directed away from the patient.
347
A
B
C
348
CAUTION
Always position the sensor windows of the airway
adapter vertically to prevent patient secretions
from obscuring the adapter windows.
C
E
D
F