Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
57
Operating concept
Loss of power
A loss of power has the following effect:
– The Cockpit switches to battery charge for up to
five minutes before performing a safe shutdown
that preserves the integrity of the patient data
and reverts to the user-defined patient default
profile.
– A medium-priority alarm is triggered at the
Cockpit and the message
please plug in
power supply
appears.
– The M540 switches to battery charge for up to
three hours before performing a safe shutdown
that preserves the integrity of the patient data
and the user settings.
Locked options
The IACS supports several locked options. For a
list of options, refer to the instructions for use
Infinity Acute Care System – Monitoring
Accessories
.
Temporary options
Temporary options make it possible for an M540 in
an IACS configuration to perform the intended
functions together with the Cockpit when the
devices do not share the same option setup. For
example, when an M540 with permanent options
docks to an IACS Cockpit that does not have the
same options activated, the M540 options
temporarily loans these options to the Cockpit.
Temporary options are deactivated when a patient
is discharged. However, they are retained if the
user turns the Cockpit or the M540 off and on.