Troubleshooting
532
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
None
(Any
SpO2
MCable)
SpO2 searching
Searching
Parameter values
are replaced by ***
The sensor is
searching for valid
pulses to compute a
measurement value.
Verify proper
sensor application.
!
(Any
Masimo
MCable)
SpO2 sensor
expired
3)
Replace sensor
Parameter values
are replaced by ***
– SpO
2
sensor
expired
– Adhesive sensor
expired
Replace the
sensor.
!
(Any
Masimo
MCable)
SpO2 sensor expires
soon
Sensor expires
soon
Parameter values
– SpO
2
sensor near
expiration
– Adhesive sensor
near expiration
Replace the
sensor.
!
(Any
SpO2
MCable)
%0 sensor failure
1)
Sensor failure
Parameter values
are replaced by ***
– Hardware failure
– Faulty SpO2
sensor
– Make sure the
SpO2 sensor is
properly
attached to the
patient and all
cables are
properly
connected.
– Replace the
sensor.
– Contact
Dräger-
authorized
service
personnel.
!
SpO2 sensor off
2)
Sensor off
Parameter values
are replaced by ***
The Masimo MCable
has detected that the
SpO2 sensor is no
longer attached to the
patient.
Reattach the SpO2
sensor.
Priority
Message
Parameter field
Problem
Solution