Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
243
Impedance respiratory rate (RRi)
Connecting the lead wires for neonatal monitoring
The ECG lead sets connect directly to the M540.
A
ECG connector on the M540
B
ECG adapter cable
C
Port cover
D
Neonatal ECG adapter cable
E
Neonatal ECG lead wires
To connect the ECG lead set
1
Insert the ECG adapter cable (B) into the
recessed ECG connector
(A) on the side of the
M540.
Orient the neonatal ECG adapter cable (B) so
the exposed pins face towards you as you push
them firmly into the ECG channel.
2
Insert the port cover (C) to protect the unused
ECG lead pins on the M540.
3
Connect the individual neonatal ECG lead wires
(E) to the neonatal ECG adapter cable (D).
For information on applying the electrodes to the
patient, refer to the figures starting on page 208.
313
D
E
B
C
A
NOTE
An ECG lead set can rest in the ECG port of the
M540 without actually being connected. Make
sure that all lead sets are pushed firmly into the
ECG port of the M540.