![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 88](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773088.webp)
Operating concept
88
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Privacy mode
Privacy mode is possible only when the patient is
admitted at the Infinity CentralStation (ICS). In
privacy mode, patient monitoring continues but the
patient data are removed from the Cockpit and the
M540 and only appears on the ICS.
Selecting privacy mode on the Cockpit
automatically activates privacy mode on the M540
and vice versa. Likewise, taking a patient out of
privacy mode on one device does the same at the
other device. Privacy mode is canceled
automatically when the connection to the Infinity
network is disrupted.
Activating privacy mode has the following effect:
– All patient data are removed from the Cockpit
and the M540 displays, but continue to be
displayed at the ICS.
– The alarm bar is deactivated.
– Acoustic alarm signals are only provided at the
ICS.
– The message
Privacy Touch Screen to
resume monitoring
is displayed in the center
of the screen.
To place the Cockpit into privacy mode
1
Select the symbol
next to the
Start/Standby...
button on the main menu bar.
2
Select
Privacy
on the toolbar.
or
1
Select the
Start/Standby...
button on the main
menu bar.
2
Select the
Start
tab if it is not already selected.
3
Select
Privacy
next to the menu selection
Display
.
The message
Privacy Touch Screen to resume
monitoring
is displayed in the center of the Cockpit
screen.
To take the Cockpit out of privacy mode
Touch the screen to activate the display of the
patient data.
NOTE
If configured to appear on the main menu bar, the
Privacy
button is also accessible on the main
menu bar. For more information, see page 452.