Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
473
System configuration
Biomed setup
–
patient monitor setup
The following table lists the available settings of the
Patient monitor
page. To access this page, see
page 470.
Selection
Available settings
Description
Change clinical
password
Use the keypad to enter the
new password (up to 4
numbers)
Configures a new password for the Cockpit.
When an M540 whose password has been
changed docks to a Cockpit with a different
password, the Cockpit password overrides the
M540 password.
CAUTION
:
Be sure to record the new
password because it cannot be retrieved
once it is lost. For further assistance,
contact DrägerService.
Change
biomedical
password
French NFC mode
– On
–
Off
(default)
When this feature is activated, HR alarms
cannot be deactivated, and the all alarm pause
period cannot exceed 3 minutes.
Simulation (basic)
– On
–
Off
(default)
Activates or deactivates basic simulation
mode.
This feature is used when the M540 is
connected. When activated, the Cockpit uses
the simulator mode from the M540 and adds
additional device connectivity parameters.
External display
– Analog
–
Digital
(default)
Selects the output for the external display.
Patient profile
selection
– On
–
Off
(default)
When this feature is activated, you can select
a profile and patient category on the
Start
dialog.
Set OR alarms
– Auto
– Manual
Auto
– OR alarms are automatically activated
when an anesthesia machine is connected.
Manual –
OR alarms must be activated
manually when an anesthesia machine is
connected.
Whenever you disconnect the A500 from the
Cockpit, the
OR Alarms
and
Cardiac bypass
features are disabled automatically regardless
of their setting.