Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
491
System configuration
To import patient profiles from a USB flash
drive
1
Insert the USB device in the USB port of the
Cockpit where you wish to transfer the profiles
to.
2
Select the
System setup...
button from the
main menu bar.
3
Select the
Biomed
tab.
4
Enter the Biomed password.
5
Select the
Service
tab.
6
Select the
Import patient profile
button to
import all patient profiles from the USB flash
drive to the Cockpit.
Messages appear in the text window informing you
if the patient profiles imported successfully or not.
To import shared system profiles from a USB
flash drive
1
Insert the USB device in the USB port of the
Cockpit where you wish to transfer the profiles
to.
2
Select the
System setup...
button from the
main menu bar.
3
Select the
Biomed
tab.
4
Enter the Biomed password.
5
Select the
Service
tab.
6
Select the
Import shared system profile
button to import all shared system profiles from
the USB device to the Cockpit. A confirmation
popup appears stating that the Cockpit will
reboot if you press the
Import
button.
7
Select the
Import
button in the confirmation
popup to start importing the patient profiles,
or
Select
Cancel
to stop the procedure and
dismiss the popup.
Messages appear in the text window informing you
if the shared system profiles imported successfully
or not. Once the system profiles are imported, the
Cockpit reboots.