Alarms
138
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Auto setting all alarm limits
The auto set function allows you to adjust alarm
limits quickly based on preset percentages listed in
the following table.
You can either auto set:
– individual parameters (see page 139)
– all parameters (see page 139)
– all ST parameters (see page 139)
Parameter
Upper limit
Lower limit
Ta, Tb, T1a, T1b,
Tblood
≤
107% of current value
≤
93%
of current value
ΔΤ,
∆
T1, PVC/min,
inCO2
Not affected
Not affected
SpO2
Adult/pediatric:
100% saturation
Neonate:
98% saturation
Current value –((value)*(5%))
ST
Current value +2.0 mm
Current value –2.0 mm
FiO
2
(via Scio)
100%
21%
etCO2
(
via Scio)
100%
18%
inHal
etHal
inIso
etIso
inEnf
etEnf
inSev
etSev
inDes
etDes
5% (+/- 0.1) above parameter value
5% (+/- 0.1) below parameter value
etCO2
Current value +25%
Current value –20%
RRc
and all others
Alarm limit that is closest to but not
more than 25% above the current value
of the parameter.
Alarm limit that is closest to but not
more than 20% below the current value
of the parameter.
NOTE
If the Auto set function forces the alarm limits of a parameter outside the allowable limit range of the
monitor, the alarm limits remain unchanged.