![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 139](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773139.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
139
Alarms
To auto set an individual parameter
1
Select the parameter field of the desired param-
eter.
2
Select the
Auto set
button located in the upper-
right corner of each parameter setup page.
or
1
Select
Sensor parameters...
on the main menu
bar.
2
Select the tab of the desired parameter.
3
Select the
Auto set
button located in the upper
right corner of each parameter setup page.
To auto set all parameters
Select the
symbol next to the
Alarms...
button
on the main menu bar >
Auto set all
.
or
1
Select the
Alarms...
button on the main menu
bar
2
Select the
Limits
tab
.
3
Select the
Auto set
button which is located in
the lower right corner below the parameter
setup table.
To auto set all ST parameters
1
Select the
Alarms...
button on the main menu
bar
2
Select the
Limits
tab.
3
Select the
ST
side tab to access the
ST
page.
4
Select the
Auto set
button which is located
below the ST table.
Alarm history and stored events
The alarm history is an electronic record of alarms
and events. The alarm history records an entry
under the following circumstances:
– An alarm occurs for a parameter whose archive
function is set to
Store
or
Str/Rec
.
These alarm
events are marked with the
symbol and
can be viewed in greater detail (see page 141)
.
– Whenever an arrhythmia event occurs (even
when the alarm function is deactivated). Only
the archive function has to be set to
Store
or
Str/Rec
.
– You
select the
Mark event
button from the main
menu bar. These alarm events are also marked
with the
symbol and can be viewed in
greater detail (see page 141)
.
– You pause alarms using the
All alarms
paused/All alarms off
button (see page 121).
– You activate Cardiac bypass mode (see
page 119).
– You activate Standby mode (see page 87).
– You select a different patient category (see
page 102).
– You transfer a patient.
– You audio pause an alarm.
– Whenever a maintenance restart event occurs.
The alarm history stores up to 150 events. When
the storage capacity of 150 events is reached, new
events replace the oldest events.
If a higher priority alarm occurs less than 5 seconds
from a previous alarm, the higher priority alarm
event is stored while the previous one is deleted.