Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
87
Operating concept
Standby mode
Temporarily interrupt patient monitoring by placing
the Cockpit and the M540 in standby mode.
Selecting standby mode on the Cockpit
automatically activates standby mode on the M540
and vice versa. Likewise, taking a patient out of
standby on one device does the same at the other
device.
Standby mode has the following effect:
– All patient data are removed from display on the
Cockpit and the M540.
– All monitoring (including acoustic and optical
alarm signals) are suppressed.
– Active alarms are considered acknowledged.
– All recordings are canceled.
– The message
Standby
–
Touch Screen to
resume monitoring
is displayed in the center
of the screen.
To place the Cockpit in standby mode
1
Select the
symbol next to the
Start/Standby...
button on the main menu bar
to display the Standby toolbar.
2
Select the
Standby
button on the toolbar.
or
1
Select the
Start/Standby...
button on the main
menu bar to display the Standby dialog.
2
Select the
Start
tab if it is not already selected.
3
Select the
Standby
button next to the menu
selection
Monitor
.
The message
Standby
–
Touch Screen to
resume monitoring
appears in the center of the
Cockpit screen.
To take the Cockpit out of standby mode
Touch the screen to resume monitoring the vital
signs of the patient.
NOTE
If configured to appear on the main menu bar, the
Standby
button is also accessible on the main
menu bar. For more information, see page 452.