Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
211
ECG, arrhythmia, and ST segment
12-lead monitoring
Standard 12-lead monitoring is only available when
you use a 6-wire lead set and a 4-wire lead set. 12-
lead monitoring using a 10-wire lead set is a locked
option that must be purchased separately. Place
the chest electrodes in positions 1 through 6 as
shown on page 210.
TruST 12-lead monitoring offers real-time
assessment of ST segment deviations with only six
electrodes. TruST uses the conventional 6-lead
standard electrode placement (see page 209),
measuring 8 leads and interpolating 4 chest leads.
TruST is available for adult and pediatric patients,
but not for neonatal patients.
You can view all ECG waveforms, including TruST,
on the
Show all
page (see page 215). For
information on how to activate TruST, see
page 234.
Accessing the ECG
functions
Select the heart rate parameter field to select
the
ECG
page directly.
or
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the
ECG
tab to access the
ECG
page.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog:
symbol and the display filter
button.
3
Select the
Settings 1, Settings 2, Show all
tabs.
The top portion of the page contains the
Auto set
and
Alarm
buttons for configuring the alarm
functions. For detailed alarm setup information, see
“Configuring the alarm settings for a patient“on
page 125.
WARNING
Do not select TruST leads for ECG signal
processing. If the QRS morphology of a TruST
lead differs from that of its equivalent
conventional lead, always refer to the
conventional lead.