![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 112](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773112.webp)
Alarms
112
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Optical alarm signals
Each alarm priority has its own distinct optical alarm
signals. When the M540 is docked on the M500,
only the Cockpit provides acoustic alarm signals.
However, optical alarm signals appear on the
Cockpit and the M540.
The alarm message in the header bar is the only
optical alarm signal if an alarming parameter is not
included in the current screen view or the alarm bar
is deactivated.
Alarm priority
Parameter field Alarm message
field
1)
in header
bar
Alarm bar
(if activated, see
page 460)
Alarm
message in
header bar
(refer to
"Messages"
on page 513
High (life-threatening)
(for example, asys-
tole, ventricular
fibrillation)
Flashing red
background
Red background
Flashing red
White alarm
message on
red
background
Medium (serious)
(for example, alarm
limit violations)
Flashing yellow
background
Yellow background
Flashing yellow
Black alarm
message on
yellow
background
Low (advisory)
(for example, dis-
connected elec-
trode)
Solid cyan
background
Cyan background
No optical alarm
signal
Black alarm
message on
cyan
background
NOTE
1)
Cockpit alarm messages are designed to be legible from a distance of 1 meter (3.3 feet) to 2 meters
(6.6 feet). M540 alarm messages are legible at arm’s length.