Troubleshooting
548
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
CO
2
monitoring
Priority
Message
Cause
Remedy
!
CO
2
incompatible
pod
The M540 has detected that the
Microstream MCable is not
compatible with the M540.
Contact DrägerService
®
.
!!
CO
2
out of range
high
1)
The parameter signal is outside
the measuring range of the
monitor.
– Check the patient and treat if
necessary.
– Check the equipment and
replace if necessary.
!
CO
2
sensor
unplugged
1)
The Microstream MCable
is
disconnected.
Check the CO
2
connections.
!
CO
2
MCable
unplugged
The Microstream MCable is
disconnected from the monitor.
Reconnect the Microstream MCable
to the monitor.
!
CO
2
MCable
failure
The Microstream MCable
hardware has failed due to one
of the following issues:
– a corrupt EPROM (erasable
programmable read-only
memory) chip
– a compromised flow rate
that caused the auto-zero
procedure to fail.
Contact DrägerService
®
.
!!!
%0 MCable: Gas
outlet blocked
The Microstream MCable gas
outlet is blocked.
Contact DrägerService
®
.