Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
159
Alarms
inDes
Increment of
0.1 kPa or 0.1%
On
Upper: 0.1 to 20.0 kPa
0.1 to 20.0%
Lower: 0.0 to 19.9 kPa
0.0 to 19.9%
12.5 kPa, 12.5%
(adult)
14.5 kPa, 14.5%
(pediatric)
0.0 kPa, 0.0%
Store
etDes
Increment of
0.1 kPa or 0.1%
Upper: 0.1 to 20.0 kPa
0.1 to 20.0%
Lower: 0.0 to 19.9 kPa
0.0 to 19.9%
20 kPa, 20%
0.0 kPa, 0.0%
Off
xMAC high
On
The user can not configure
the xMAC limits.
Not applicable
Not applicable
Store
Parameter Alarm
default
status
Alarm limit range
Upper limit
defaults
Lower limit
defaults
Archive
default
setting