Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
203
ECG, arrhythmia, and ST segment
Connecting the lead wires for neonatal monitoring
The ECG lead sets connect directly to the M540.
A
M540 ECG port
B
Neonatal ECG adapter cable
C
Neonatal ECG electrodes
D
Port cover
To connect the ECG lead set
1
Insert the neonatal ECG adapter cable (B) into
the ECG port (A) on the side of the M540.
Orient the neonatal ECG adapter cable (B) so
the exposed pins face towards you as you push
them firmly into the ECG port.
2
Insert the port cover (D) to protect the unused
ECG lead pins on the M540.
3
Connect the individual neonatal ECG lead wire
(C) to the neonatal ECG adapter cable (B).
For information on applying the electrodes to
the patient, refer to the figures starting on
page 208.
313
A
B
C
D
NOTE
An ECG lead set can rest in the ECG port of the
M540 without actually being connected. Make
sure that all ECG lead sets are pushed firmly into
the ECG port of the M540.