Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
381
External Device – Bispectral index (BIS)
BIS precautions
Device compatibility
The Cockpit supports the Aspect BIS Vista
Complete 2-channel monitor with software version
3.00.
BIS display
On the Cockpit, the BIS x display consists of:
– BISx parameter field
– 1 EEG waveform
BIS parameter field
BIS parameter fields report parameter values. For
detailed information regarding the content of
parameter fields for each parameter, see the
chapter "Troubleshooting" on page 509.
WARNING
When connecting a third-party device, verify
its proper operation before clinical use. Refer
to the instructions for use of the third-party
device for complete instructions. For further
questions, contact your local representative.
WARNING
Always refer to the primary data source before
making diagnostic or therapeutic decisions.
Connecting peripheral devices is supported
via the device connectivity option. Only
connect peripheral medical devices to a
patient monitor if those devices comply with
the electrical safety requirements found in IEC
60601-1. Refer to "Safe connection with other
electrical equipment" on page 13 of these
instructions for use for information on how to
connect devices safely.
WARNING
The following section lists all of the external
devices and related software versions that
Dräger has validated. Dräger cannot make any
claim for the reliability of the data for
subsequent or previous software versions or
for any devices that have not been validated.
In the interest of patient safety and device
performance, do not connect devices to the
monitor which have not been approved by
Dräger. The hospital is responsible for
contacting Dräger to determine the
compatibility and warranty status of any
connection made to another manufacturer’s
medical devices.
NOTE
The following diagram shows a typical parameter
field layout. This layout may change when addi-
tional parameters are put on display. For more in-
formation, see "Parameter fields" on page 62.