Troubleshooting
510
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Overview
Alarm messages in the alarm display field are
displayed in hierarchical order.
For example, if two faults are detected
simultaneously, the more urgent of the two is
displayed.
The priority level of the alarm messages (see
page 109 for definitions) is identified by
exclamation marks:
If no priority level is assigned, the message is
informational and no action is required.
In the following table, messages are listed in
alphabetical order. These tables identify possible
alarm causes and provides corrective action. The
various causes and remedies should be worked
through in the order listed until the problem has
been resolved.
Device communication messages
Warning =
!!! Message of high priority
Caution =
!! Message of medium priority
Advisory = !
Message of low priority
Priority
Message
Cause
Remedy
None
Check IP address
Duplicate PDS or Multicast
address.
Ensure the fourth octet of the IP
address is unique within the mon-
itoring unit.
None
Disconnected from
M540
The M540 is disconnected
from the M500.
Dock the M540.
None
Duplicate device
name
Duplicate domain.
Assign a unique domain name.
None
Duplicate IP address
The IP address is already in
use.
Assign a unique IP address.