![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 298](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773298.webp)
Non-invasive blood pressure (NIBP)
298
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Continuous measurements
In continuous mode, the M540 continuously
initiates NIBP measurements over a 5-minute
period.
A 10 second (±1 second) minimum interval
between the end of one measurement and the start
of another provides minimal perfusion of the limb.
To activate or deactivate continuous mode
1
Select the non-invasive blood pressure
parameter field to select the
NIBP
page directly,
or
2
Select
Sensor parameters...
from the main
menu bar.
3
Select the
NIBP
tab to access the
NIBP
page.
4
Select
On
or
Off
next to
Continuous mode
.
To stop continuous measurements
Press the
NIBP start/stop
button on the main
menu bar,
or
Deactivate
Continuous mode
in the
NIBP
page (see page 300).
WARNING
When using continuous mode, observe the
patient closely and verify limb perfusion
clinically. Be extra vigilant when using
continuous mode on neonates or
hemodynamically compromised patients.
NOTE
Continuous non-invasive blood pressure mode
prevents you from enabling venous stasis.