Troubleshooting
518
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
!
%0 Unplugged
1)
%0 leads off
1)
The parameter value
is replaced by ***
Lead-off condition
detected due to:
– broken
cable(s)
– disconnected
ECG lead
wires
– loose lead
wire(s)
– wrong lead
selected
– dried out
electrode gel
ECG cable(s)
disconnected from
the M540.
– Replace faulty
cable(s).
– Reapply gel on
reusable
electrodes and
reapply them or
replace new
disposable
electrodes.
– Select another
ECG lead for
processing.
– If monitoring
augmented leads,
verify that the
number of
selected leads in
the ECG setup
page is correct.
– Check cable(s)
and connection(s)
– Replace cable(s)
if necessary
None
None
None
None
None
None
LA lead off
LL lead off
RA lead off
RL lead off
V lead off
V1 lead off
V2 lead off
V3 lead off
V4 lead off
V5 lead off
V6 lead off
V+ lead off
Parameter value
The indicated lead
is no longer
attached to the
patient.
Reattach the
electrode to the
patient.
1)
%0 is a placeholder for the parameter label HR or ECG.
Priority
Message
Parameter field
Problem
Solution