Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
469
System configuration
Rest
ECG
setup
Appropriate settings are crucial for optimal 12-lead
analysis. The following table lists the available
settings of the
Rest ECG report
page which
controls various settings. To access this page, see
page 467.
To obtain an optimal automatic diagnostic
interpretation of an Rest ECG report, make sure the
selections in the following table and the
Weight
,
Height
,
and
Birth date
in the
Demographics
page
(see page 101) are configured appropriately for the
patient.
NOTE
The Rest ECG report is only available for adult
and pediatric patients.
NOTE
If the ECG Filter is set to
ESU
, Rest ECG report
cannot be generated.
Selection Available
settings
Description
Gender
–
Unknown
(default)
– Male
– Female
The selected information is included in the
report.
Race
–
Unknown
(default)
– Caucasian
– Asian
– African
– Other
Medication 1
No meds
,
Unknown
(default),
list of medications
Medication 2
Condition 1
Pick list with several choices
for indicating the medical
condition of the patient.
(
Unknown
is the default)
Condition 2
Notes
List of entries for annotating
the condition of the patient.
(
None
is the default)
Rest ECG report
button
The button is grayed out and not selectable
when:
– The patient is not admitted at the ICS.
– The Rest ECG analysis feature is not
activated at the ICS.
– The 12-lead ECG option is not unlocked.
– The required 12-lead cable is not
connected.