Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
359
Scio Monitoring
The gas analyzer is available in four variants with
different functions as listed below.
IACS automatically detects the variant of gas
analyzer connected and adjusts all context-
sensitive menus for the gas analyzer variant.
At start-up, the gas analyzer warms up and displays
the low-priority alarm
Scio warming up: Accur.
low
on the M540 monitor. During this time,
concentrations for certain gases may not be
available and the anesthetic agent may not be
identified.
The following alarms are supported prior to the de-
tection of a valid breath:
–
Mixed Agents
–
%0 out of range high
–
%0 reduced accuracy
–
%0 sensor failure
–
%0 value temporarily unavail.
–
Check water trap/sample line
–
Gas sensor failure
–
CO
2
reduced accuracy
–
CO
2
sensor failure
– Gas sensor reduced accuracy
– <agent>>%val
– inN
2
O >82
– FiO
2
<%val
– Inspiratory xMAC high
–
Sample line blocked
–
Scio is not connected
–
Scio unavailable for neonates
–
Scio warming up: Accur. low
–
Second agent detected
–
Water trap is full
All other
O
2
alarms, CO
2
alarms,
N
2
O
alarms, and
anesthetic gas alarms are active only after one
breath has been detected.
O
2
CO
2
, N
2
O
Agent
Agent ID
Mixtures
Scio Four
No
Yes
1 out of 5
No
No
Scio Four Oxi
Yes
Yes
1 out of 5
No
No
Scio Four plus
No
Yes
2 out of 5
Yes
Yes
Scio Four Oxi plus
Yes
Yes
2 out of 5
Yes
Yes
NOTE
The "
%0
" symbol indicates CO
2
, N
2
O, O
2
, and
any agents that may appear in the alarm
message.