Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
515
Troubleshooting
!!!
HR, ASY, VF off
This message appears in the alarm
message field under the following
circumstances:
– Heart rate alarms are
deactivated,
–
ASY/VF alarms
feature is
set to
Follow HR alarm
(see
page 460),
– Arrhythmia monitoring is
deactivated.
The same message also appears
under the following circumstances:
– Heart rate alarms are
deactivated,
–
ASY/VF alarms
feature is
set to
Always on
(see
page 460),
– Arrhythmia monitoring is
deactivated,
– The selected
HR source
is
activated and is either
SpO2
or
ART
.
Activate the functions to
remove the message.
None
Pacer off
Pacer fusion
These messages appear above the
ECG waveform when the
corresponding function is activated
or deactivated (see page 214)
Deactivate the function to
remove the message.
None
Privacy
Touch Screen to
resume monitoring
This message appears in the center
of the Cockpit screen when privacy
mode has been activated
(see page 88). All patient data are
removed from the screen and are
only visible at the ICS. This function
is not available unless the patient is
also admitted at the ICS.
Take the patient out of
standby to view all the data
at the Cockpit.
None
Standby
Touch Screen to
resume monitoring
This message appears in the center
of the Cockpit screen when the
Cockpit has been placed in standby
mode.
Touch the screen to resume
monitoring.
Priority
Message
Cause
Remedy