Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
137
Alarms
Configuring ST alarm settings
Some of the ST alarm settings described are also
available on the
ST alarms
page (see page 233). In
the following steps, the letters in parentheses refer
to the diagram of the
Limits
>
ST
page (see
page 135). Alarm ranges and defaults are listed
starting on page 149.
To change ST alarm settings
1
Select the
Alarms...
button on the main menu
bar.
2
Select the
Limits
tab (if not already selected).
3
Select the
ST
tab (H) on the right side of the
page.
4
Select the setup button in the
Alarm
column (C)
to activate or deactivate alarm monitoring. A
crossed-out triangle appears in the parameter
field when alarm monitoring is deactivated.
5
Select the corresponding setup button in the
Lower
column (D) to adjust the lower alarm
limits.
6
Select the corresponding setup button in the
Upper
column (F) to adjust the upper alarm
limits.
7
Use one of the following settings in the
Archive
column (G) to determine what happens in
response to an alarm:
–
Off
– no event is stored and no recording is
generated.
–
Store
– stores the event for later review
(see page 140).
–
Record
– generates a timed recording
–
Str/Rec
– generates a timed recording and
stores the event.
8
Use the
Auto set all
(I) button to adjust the
alarm limits for all ST parameters (see
page 139).
9
Use the
Event duration [s]
button (J) to select
a time an upper ST alarm limit has to be in
violation before an alarm is triggered (see
page 111).