Maintenance
570
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Inspection / safety checks
Inspection and safety checks of devices must be
performed according to the suggested intervals
specified in the table on page 569.
Scope of inspection/safety checks for
the Cockpit (C500/C700)
The safety checks are no substitute for preventive
maintenance measures (including preventive
replacement of wearing parts) as identified by the
manufacturer.
1
Check accompanying documents:
– Instructions for use are available
2
Perform a functional test of the following
features according to the instructions for use:
– Verify the LEDs
– Perform system tests
3
Check that the device combination is in good
condition:
– All labels are complete and legible
– There is no visible damage
– Fuses which are accessible from the
outside are in compliance with the specified
values
4
Use the instructions for use to check that all
components and accessories needed to use the
product are available.
5
Check the electrical safety requirements
according to IEC62353.
6
Verify that the optical and acoustic alarm
signals function properly.
Scope of inspection/safety checks for
the PS250 / P2500
The safety checks are no substitute for preventive
measures (including preventive replacement of
wearing parts) as identified by the manufacturer.
1
Check accompanying documents:
– Instructions for use are available
2
Perform a functional test of the following
features:
– Verify the LEDs
– Perform system tests
3
Check that the device combination is in good
condition:
– All labels are complete and legible.
– There is no visible damage.
– Fuses which are accessible from the
outside comply with the specified values.
4
Check the electrical safety requirements
according to IEC62353 every two yea rs by
qualified DrägerService personnel.
5
Check the following safety features:
– The power LED and the battery indicator
LED function properly.
– The C500/C700 are powered correctly.
– Check the functional integrity of the Infinity
MCable – Nurse call.
WARNING
Risk of medical device failure
If safety checks are not performed on a
regular basis, the proper operation of the
medical device can be compromised.
Perform safety checks at the indicated
intervals.