Troubleshooting
530
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
!
(Any
SpO2
MCable)
SpO2 H/W failure
Parameter values
are replaced by ***
Hardware failure
– Check for faulty
MCable
– Contact
Dräger-
authorized
service
personnel.
!
SpO2 Interference
Detected
2)
Interference
detected
Parameter values
are replaced by ***
Interference such as
artifact or too much
ambient light was
detected.
– Make sure the
sensor is
properly
attached.
– Make sure that
no nail polish or
some other
substance is
blocking the
light.
– Change the
sensor location.
1)
%0 is a placeholder for the parameter label SpO2.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
3)
In the parameter field the parameter value is replaced by ***
Priority
Message
Parameter field
Problem
Solution