Alarms
118
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Zeroing invasive blood pressures
Zeroing all invasive blood pressures from the
Zero
all
button on the Cockpit menu bar or from the
key on the hemodynamic pods (see page 314) has
the following effects:
– All invasive blood pressure and CPP limit
alarms and static alarms are suppressed from
the time the button/key is pressed until
30 seconds after the zeroing procedure is
completed. This includes an alarm for a
disconnected arterial catheter.
Zeroing an individual blood pressure from a specific
invasive blood pressure page on the Cockpit (see
page 313) has the following effects:
– The invasive blood pressure limit alarm for that
parameter is suppressed from the time the
button is pressed until 30 seconds after the
zeroing procedure is completed.
– If the zeroed parameter is ICP or ART, the CPP
limit alarm is also suppressed from the time the
button is pressed until 30 seconds after the
zeroing procedure is completed.
The following alarm conditions cancel the
suppression of alarms caused by zeroing invasive
blood pressures:
– The invasive blood pressure parameter is
outside (high/low) the measuring range.
– Invasive blood pressure hardware failures such
as a transducer failure
– Unplugged transducers
– Disconnected hemodynamic pods
– A wedge pressure measurement that ends
before the 30-second zeroing period ends will
activate the alarm limit for the parameter PA M
only
Privacy mode
When privacy mode is activated, the following
happens at the Cockpit:
– All patient data are removed from the Cockpit
and the M540 but continue to be visible at the
ICS (Infinity CentralStation).
– The Cockpit and at the M540 display the alarm
message
Privacy
Touch Screen to resume
monitoring
.
– The alarm bar is deactivated.
– Acoustic alarm signals are only provided at the
ICS.
–
Home
is the only active button on the main
menu bar of the Cockpit; all other buttons are
inactive.
You can activate privacy mode only if the patient is
also admitted at the ICS. To activate or deactivate
this feature, see page 88.
Standby mode
When Standby mode is activated, the following
happens at the Cockpit:
– All patient data are removed from the screen.
– All monitoring (including acoustic and optical
alarm signals) is suppressed.
– Active alarms are considered acknowledged by
the user.
– The message
Touch Screen to resume
monitoring
appears at the ICS, at the Cockpit,
and at the M540.
–
Home
is the only active button on the main
menu bar of the Cockpit; all other buttons are
inactive.
– All recordings are canceled.
To activate or deactivate this function, see page 87.