System configuration
464
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Alarm setup – Code functions
For urgent care, you can configure a set of
individual monitoring functions. These functions
can be activated simultaneously when you select
the
Code
button on the main menu bar. For more
information regarding this function, see page 147.
When the
Code
button is pressed, a timer along
with a
Stop
and a
Reset
button appears in the
header bar.
The following table lists the available settings of the
Code
page. To access this page, see page 458.
Audio pause: Quiet
mode
–
On
– Off
(default)
–
On
– Only alarm conditions of high priority
override an active audio pause. The
appropriate parameter field flashes. Alarm
conditions of equal or lower alarm priority
will not be reported with an alarm tone.
–
Off
– Any new alarm condition, regardless
of its alarm priority, overrides an already
active audio pause state at the Cockpit and
at the ICS if the patient is admitted there. All
optical and acoustic alarm signals are
reported fully for any new alarm condition.
For detailed information how quiet mode affects
the audio pause behavior, see page 121.
Selection
Available settings
Description
Selection
Available settings
Description
Continuous
recording
– On
–
Off
(default)
–
On
– a continuous recording starts when
you select the
Code
button.
–
Off
– no recording starts when you select
the
Code
button.
Continuous NIBP
mode
– On
–
Off
(default)
–
On
– continuous NIBP measurements start
when you select the
Code
button.
–
Off
– no NIBP measurements start when
you select the
Code
button.
Audio off
– On
–
Off
(default)
–
On
– No acoustic alarm signal sounds when
you select the
Code
button.
–
Off
– Alarm tones sound when when you
select the
Code
button.