![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 49](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773049.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
49
Operating concept
Network data transfer
The IACS supports the transfer of patient data to
and from the following devices assigned to the
same monitoring unit:
– Infinity Delta/Delta XL/Kappa (software version
VF7 to VF9.x)
– Other IACS monitoring Cockpits
The following data are included in a patient data
transfer:
– Demographic data (see page 101 for
information of what demographic data is
included)
– Trends (up to 60 trended
parameters)
– Events containing up to 32 parameters for
(C500) or up to 40 parameters for (C700)
– Hemodynamic, oxygenation, and ventilation
calculation results
– Laboratory data values are not transferred
during a network transfer. The following
diagram shows the
Transfer
page which is
used for patient data transfers.
A
Transfer
tab
B
Care area
selection arrow
C
Current patient
column
D
Device name
column
E
Start transfer
button
To transfer data over the network
The IACS network supports the transfer of patient
data from a source device, such as an Infinity Delta,
Delta XL/Kappa or an IACS Cockpit, to another
Cockpit.
1
Place the source device in standby mode.
2
Go to the Cockpit to which you wish to transfer
data.
NOTE
You can also transfer patient data by undocking
and redocking an M540.
NOTE
The amount of data being transferred over the
network depends on how much data is available
at the source device.
A maximum of 60 trend
parameters can be transferred over the network
based on the parameter priority of the Cockpit.
For Delta series monitors, the maximum amount
of data is 24 hours. For a network transfer
between IACS Cockpits, the maximum of data is
96 hours.
NOTE
Due to a possible time variation between cockpits,
the trend time stamp may vary by one minute after
a network transfer.
006
A
C
D
B
E